Sensor Calibration

We provide calibration tools for new GelSight sensors, following the method described in [1]; the process generally takes less than an hour. For lab-made sensors, refer to examples/configs/gsmini.yaml to create a sensor specification, which will be used as the CONFIG_PATH in the command below. Below are step-by-step instructions for sensor calibration. Note: The current version does not support gel pads with markers.

Calibration Data Collection

Ball Indenter Collecting Data Collected Image

To collect calibration data, use a ball indenter of known diameter to press against the sensor. Examples of the setup and resulting images are shown above. Run the following command and allocate a location CALIB_DIR to save the calibration data:

collect_data [-b CALIB_DIR] [-d BALL_DIAMETER_IN_MM] [-c CONFIG_PATH]
  • Instruction:
    • Save background image: Press 'b'
    • Capture ~50 tactile images by pressing the ball in various locations: Press 'w'
    • Exit: Press 'q'
  • Tips for Optimal Calibration:
    • Ball Size: Select a ball that appears well-sized within the sensor’s view, like the tactile image shown above; 4mm to 9mm is suitable for GelSight Mini.
    • Pressure: Avoid pressing too hard.
    • Coverage: Capture at least 50 images with the ball contacting different regions of the sensor.
    • Using Multiple Balls: Use the same CALIB_DIR and specify distinct BALL_DIAMETER values if balls in different size are applied.

Label Collected Data

NanoGui Screenshot

Run the command below to label the contact circle on the collected tactile data using NanoGUI:

label_data [-b CALIB_DIR] [-c CONFIG_PATH]
  • Instruction:
    • Click the Open icon to begin.
    • Keyboard Controls for aligning the label with the contact circle:
      • Arrow keys (left/right/up/down): Adjust the circle's position.
      • 'm' / 'p': Decrease / increase the circle's radius.
      • 'f' / 'c': Decrease / increase the circle's movement step.
    • Once aligned, click the Calibrate icon.
    • After labeling all data, close the window to exit.

Prepare Dataset

Run the command below to prepare the dataset for calibration model training:

prepare_data [-b CALIB_DIR] [-c CONFIG_PATH]

Train Calibration Model

Train the MLP model to map pixel color and location (RGBXY) to surface gradients for each pixel. Use the following command to train the model with the collected dataset:

train_model [-b CALIB_DIR] [-d {cpu|cuda}]

The trained model is saved in CALIB_DIR/model/nnmodel.pth.

Test the Trained Calibration Model

Once the model is trained, connect the sensor and run the following command to stream images and perform real-time surface reconstruction using the trained calibration model:

test_model [-b CALIB_DIR] [-c CONFIG_PATH]

After starting, wait briefly for background data collection; real-time surface gradient predictions will then be displayed. Press any key to exit.

References

  1. S. Wang, Y. She, B. Romero, and E. H. Adelson, “Gelsight wedge: Measuring high-resolution 3d contact geometry with a compact robot finger,” in 2021 IEEE International Conference on Robotics and Automation (ICRA). IEEE, 2021.